1. 韦德国际1946(bevictor)

      Fluorescent image acquisition and processing using Axiovert 200M microscope and ImageJ software

      Electra Brunialti, Alessandro Maria Villa, Paolo Ciana

      Published: 2022-11-24 DOI: 10.17504/protocols.io.ewov1o98plr2/v1
      AI 解读
      浏览 1527

      Abstract

      Fluorescent image acquisition and processing using Axiovert 200M microscope and ImageJ software to analyze morphological and dynamic changes of primary fluorescent microglia.

      Steps

      Step 1: Acquire live microglia images using Axiovert 200M microscope with dedicated software (AxioVision Rel 4.9, Zeiss)

      1.

      Insert the cell culture plate in the microscope holder, and set chamber parameters: T 37°C and CO2 5%.

      2.

      Using X20 magnification chose 20 random fields;

      3.

      Set exposition of the fluorescent channel to have a sharp image of microglia bodies and branches;

      4.

      Records the live fluorescent microglia for 2 h taking a picture every 5 min.

      5.

      Save the recorded file as a “.zvi”.

      Step 2: Elaborate the acquired images using Fiji software (ImageJ, NIH, version 2.0.0)

      6.

      Open ".zvi" file with Fiji software as hyperstack and tick the “split channel” option;

      7.

      close the bright field channel and start to elaborate the fluorescent channel (microglia);

      8.

      subtract the background using “Process › Subtract Background”(identifier: legacy:ij.plugin.filter.BackgroundSubtracter);

      9.

      defined threshold (foreground) that corresponds to green fluorescent objects using “Image › Adjust › Threshold” (identifier: legacy:ij.plugin.frame.ThresholdAdjuster). Keep the threshold consistent between acquisitions;

      10.

      apply despeckle function “Process › Noise › Despeckle” (identifier:legacy:ij.plugin.filter.RankFilters("despeckle"));

      11.

      apply smoothing function “Process › Smooth” (identifier: legacy:ij.plugin.filter.Filters("smooth"));

      12.

      set the measurement: “Analyze › Set Measurements” (identifier: legacy:ij.plugin.filter.Analyzer("set")), and tick “Area”, “Center of Mass”, “Feret’s Diameter” and “Shape Descriptors”;

      13.

      for each microglia select the area that contains the microglia in each time-frames using “Edit › Options › Roi Defaults” (identifier:legacy:ij.gui.RoiDefaultsDialog);

      14.

      run analyze particles “Analyze › Analyze Particles” (identifier:legacy:ij.plugin.filter.ParticleAnalyzer), set size (micron^2):130-infinity, circularity: 0.00-1.00; tick “display results”;

      15.

      process “all images”;

      16.

      copy the data in an Excell file;

      17.

      repeat the steps from 13 to 16 for each microglia.

      18.

      Among the "Shape Descriptors", keep the values of “Area”, “Solidity”, “FeretAngle”, “XM” and “YM”;

      19.

      to calculate the distance covered by the cell during the time-lapse use the coordinates of the center of mass and sum the distance covered in each time frame assuming that the distance between frames corresponds to the cathetus of a right triangle made by X-axis and Y-axis displacement;

      20.

      to calculate the rotation sum the “FeretAngle” taking into account that it is the angle among Ferret’s diameter and parallel line to the cell contour only on x-axis;

      21.

      calculate the median and the CV% of "Solidity" and "Area";

      22.

      to perform the cluster analysis use each parameter obtained from the analysis; use the values of the vehicle and treated cells and identify the median parameter for the experiment;

      23.

      use the identified median as a threshold to cluster the cells in two groups (over or under the median);

      24.

      combine two parameters to generate four different clusters.